DKSH.com.au Suppliers Contact

Stuart Melting Point Apparatus, Digital, Display resolution 1 degree

Code: SMP10

  • Description
  • Specs



  • Digital selection and display of temperature

  • Choice of accuracy, 1 degree resolution 

  • Easy to operate with plateau facility

  • Ideal for educational use

  • Supplied with calibration certificate


The SMP10 has been designed with safety and ease of operation in mind making it ideal for use in education. The temperature is selected, measured and displayed digitally making it accurate and negating the need for a thermometer. The SMP10 displays temperature to 1 degree resolution while the more advanced SMP20 has a 0.1 degree resolution as well as a variable ramp rate and hold key so that the exact melt temperature can be recorded.

Two samples can be tested simultaneously. They are viewed via a magnifying lens with clear observation aided by built in illumination. Extendible back feet allow the unit to be operated at the optimum viewing angle. Full access to the block aids cleaning. The simple to follow instructions are printed directly on the instrument in most European languages for ease of use.

To operate simply select a plateau temperature via the digital display and press "start". The unit quickly heats up and remains at the selected plateau temperature until the user is ready to start the test. Insert the sample tubes and press "start". The unit then heats at a fixed rate of 2°C per minute for the SMP10 and at a user selected rate of between 1 and 10°C per minute for the SMP20. When the sample is seen to melt, note the temperature on the display. Press "stop" to end heating and cool the block.

Supplied complete with calibration certificate showing individual serial number for traceability.

Model                                SMP10                                                               

Number of samples        2                                                                          

Temperature range      Ambient to 300ºC                                            

Temperature accuracy  ± 1.0ºC at 20ºC, ± 2.5ºC at 300ºC                

Display                               Three digit LED                                               

Display resolution           1ºC                                                                    

Display hold facility          No                                                                       

Ramp rates                    20ºC per minute to plateau, 2ºC per minute to melt                                    

Dimensions (w x d x h)  160 x 220 x 170mm                                      

Net weight                          1.8 Kg                                                           

Electrical supply                230V, 50Hz, 75W                                           

 
Contact us
  • DKSH Australia Pty. Ltd.
  • 1 Beyer Road
  • Braeside VIC 3195
  • Australia

Shop online safely and securely