CDS 9350 Thermal Desorption system
The CDS 9350 Thermal Desorption system is designed for continuous real time air monitoring with a dual thermal desorption tube setup. It features a hand-held controller for system maintenance and direct communication. With the highest programmable valve temperature on the market (350°C), the 9350 can venture into applications where other systems can not perform, including VOCs, sVOCs, polar and nonpolar compounds. The two stage trapping and desorption process with Pre-desorb and Pre-heat feature eliminate the need of any Peltier cooling or liquid cryogen to improve the resolution for early eluting molecules. An optional heated sampling line is available
9350 Specifications
-
Programmable Time's: 0 - 999.9 minutes
-
Focus Trap: Yes, Standard 1/8" Vocarb; Low Thermal Mass
-
Dynamic Range: C2-C33+
-
Cyrogen: Not Required. Standard Pre-Desorb
-
Temperature Accuracy: 1°C
-
Tube Heating Rate: 1000°C/min
-
Trap Heating Rate: 900°C/min
-
Valve Oven: Up to 350°C
-
Transfer Line: Up to 350°C
-
Sample Desorb: Up to 400°C
-
Focus Desorb: Up to 400°C
-
On-Line Sampling: Continuour with Dual Tubes
-
GC Connection: Heated Transfer Line
-
Weight: 21 kg
-
Dimension (W x H x D): 16" X 11" X 18"